Validation of a Hybrid Blood Pressure Monitor to Replace the Mercury Standard for Validation Studies According to Modified Criteria of the European Society of Hypertension International Protocol 2010
In Brief
A Phase 4 clinical trial evaluating Blood Pressure Measurement for Essential Hypertension. Completed, enrolled 51 participants across 1 site.
Detailed Summary
Assessment of the accuracy of a Blood Pressure Measuring Device with a digital display-column, NISSEI DM 3000, according to an adaptation of the "European Society of Hypertension International Protocol Revision 2010 for the Validation of Blood Pressure Measuring Devices in Adults".
Study Details
Timeline
Interventions
All patients have their blood pressure measured simultaneously by two observers using parallel connected mercury sphygmomanometers and by a supervisor using the tested device, NISSEI DM 3000. Six sets of measurements are performed.