Search/NCT02111993
NCT02111993N/ACompleted

Evaluating Myocardial Injury During ICD Implantation Using the Upper Limit of Vulnerability (ULV) Method vs. Standard Defibrillation Threshold Testing

University of California, San Diego·interventional·Posted Apr 11, 2014·Updated May 22, 2019

In Brief

A clinical study evaluating Standard Defibrillation Testing and Upper Limit of Vulnerability Testing for Cardiomyopathy and 2 related conditions. Completed, enrolled 60 participants across 1 site.

Detailed Summary

The objective of this study is to evaluate myocardial injury, if any, as quantified by cardiac markers (Troponin-T) in defibrillation threshold (DFT) testing during implantation of implantable cardioverter defibrillators (ICDs) using the upper limit of vulnerability (ULV) method vs. standard defibrillation threshold method.

Study Details

Study Typeinterventional
Allocation--
Masking--
Primary Purpose--
CountriesUnited States
CollaboratorsMedtronic

Timeline

N/ACompletedFinished
201220132014201520162017201820192020202120222023202420252026
First PostedApr 11, 2014
Enrollment StartAug 1, 2011
Primary CompletionApr 1, 2015
Study CompletionJun 1, 2015
TodayJul 2, 2026
Enrollment to primary: 3.7 yearsPosted 12.2 years ago

Interventions

Standard Defibrillation Testingdevice

Once the ICD is implanted, Standard Defibrillation Threshold (DFT) testing will begin. A low energy shock will be delivered to the heart at the peak of the T wave to induce ventricular fibrillation (VF). Once VF is induced, a rescue shock of 25J will be delivered to restore sinus rhythm. This process will be repeated once more after 5 minutes.

Upper Limit of Vulnerability Testingdevice

Once the ICD is implanted, Upper Limit of Vulnerability (ULV) testing will commence. An 18 J shock will be delivered will at different points from the t-wave. If VF is not induced with any of these shocks, then the ULV will be considered to be 18J. If VF is induced, then a 25 J rescue shock will be delivered.