Desaturation Validation of INVSENSOR00038
In Brief
A clinical study evaluating INVSENSOR00038 for Healthy. Completed, enrolled 13 participants across 1 site.
Detailed Summary
This study is designed to evaluate the trending accuracy of a noninvasive regional oximetry measurement of somatic oxygen saturation. One investigational sensor (INVSENSOR00038) will be placed on the somatic site of the volunteer. The values obtained by the test sensor will be compared to the reference value obtained from an FDA-cleared sensor placed on the somatic site on the contralateral side. Data will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's oxygen saturation is approximately 70%. The subject will then be returned to inhaling room air.
Study Details
Timeline
Interventions
Noninvasive regional oximeter