Results of the SpO2 and Pulse Rate Accuracy Comparison of VitalDetect™ to Arterial Blood CO-Oximetry and Reference ECG
In Brief
A clinical study evaluating Accuracy of the VitalDetect™ pulse oximetry system for Arterial Disease. Completed, enrolled 10 participants across 1 site.
Detailed Summary
Clinical Investigation Plan: Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry Study ID# PR 2019-353, Version: Rev 1, Date: 15 Oct 2019 Study Dates: November 4-5, 2019
Study Details
Timeline
Interventions
The purpose of this study was to validate the SpO2 accuracy of the VitalDetect™ pulse oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system would meet the required specification of ARMS of 3.0% or less.