CI

At a glance

ClinicalIndex Comparison Record
N/ACompleted· 160 enrolled
Drug / intervention
NIRAF Detection Technologydevice
Likely dose
Not stated in record
Structured eligibility isn't available for this trial yet — see the full criteria in the Eligibility tab below.

Standardized by ClinicalIndex from the ClinicalTrials.gov record · verify against the source.

Search/NCT04299425
NCT04299425N/ACompleted

Evaluating Impact of Near Infrared Autofluorescence (NIRAF) Detection for Identifying Parathyroid Glands During Parathyroidectomy

Vanderbilt-Ingram Cancer Center·interventional·Posted Mar 6, 2020·Updated May 13, 2024

In Brief

A clinical study evaluating NIRAF Detection Technology for Parathyroid Adenoma and 5 related conditions. Completed, enrolled 160 participants across 1 site.

Detailed Summary

This study describes a single center, randomized, single-blinded clinical trial to assess the clinical benefits of the use of near infrared autofluorescence (NIRAF) detection with an FDA-cleared device 'Parathyroid Eye (PTeye)' for identifying parathyroid glands (PGs) during parathyroidectomy (PTx) procedures. It compares risk-benefits and outcomes in PTx patients where NIRAF detection with PTeye for parathyroid identification is either used or not used.

Study Details

Study Typeinterventional
Allocation--
Masking--
Primary Purpose--
CountriesUnited States

Timeline

N/ACompletedFinished
202120222023202420252026
First PostedMar 6, 2020
Enrollment StartMar 13, 2020
Primary CompletionMar 3, 2023
TodayJul 2, 2026
Enrollment to primary: 3.0 yearsPosted 6.3 years ago

Interventions

NIRAF Detection Technologydevice

Near Infrared Autofluorescence (NIRAF) detection technology or 'PTeye' consists of a disposable fiber-optic probe that emits non-ionizing radiation from a NIR 785 nm laser source, and also transmits the resulting tissue NIRAF to a photo detector. The 785 nm laser source emits a maximum power of 20 mW. The device is FDA cleared for clinical use in general surgeries and dermatological use (Class 2 device). After surgeon identifies a potential parathyroid gland in the surgical field, the surgeon places the fiber-optic probe of PTeye on suspect tissue and presses the device foot-pedal to activate tissue NIRAF measurement. Auditory beep at high frequency with a Detection Ratio \> 1.2 is interpreted by device as the suspect tissue being positive for parathyroid.